GMH 3691 GL
Artikel Nr. GMH 3691 GL
Air oxygen measuring-device
- Double display for oxygen and temperature
- Meas. units O2 -concentration and O2 -partial
pressure
- Alarm detector with integrated horn
- Serial interface
- Automatic temperature compensation
- Wide range of application: e.g. bio chemistry, mediccine and food
technology,
safety at work, air conditioning and
ventilation technology, sports etc.
- Min-/Max-value memory
- Hold function
- Battery- and d.c. operation
- Extremely simple calibration in atmospheric air
-compensation of atmospheric pressure by entering the atmospheric pressure
- 1-point or 2-point-calibration
GMH3691 GL (Electrode not included -
please order separately!)
Specification:
Measuring range:
Oxygen concentration: 0.0 ... 100.0 % O2
(gaseous)
Oxygen partial pressure (GMH3691): 0 ... 1100 hPa O2
Temperature: -5.0 ... 50.0°C
Accuracy (device):
± 0.1%
±1digit (O2 -concentration)
±1 hPa ±1digit (O2 -partial pressure)
±0.2°C ±1digit (temperature)
Sensor connection:
6 pin screened
Mini-DIN socket
suitable sensor p.r.t.
electrodes (GGO 369 / GOO 369)
Display:
2 four-digit LCDs (12.4 mm and 7
mm high) for oxygen and temperature
and for min./max. value, hold
function, etc. as well as additional functional arrows
Working temperature:
0 to +50°C
Relative humidity:
0 to +95% r.h. (not
condensing)
Storage temperature:
-20 to +70 °C
Interface:
serial interface (3.5mm jack
connector),
direct connection to RS232 interface
of a PC via electrically isolated interface adapter
GRS
3100
or
GRS
3105 or connection to USB-interface via USB3100-adapter
Min-/Max-value memory:
max. and min.
values will be memorized.
Hold function:
by pressing a button the
current meas. value will be memorized.
Alarm:
integrated limit detector for
min. or max.-alarm
Push buttons:
6 membrane keys for ON/OFF
switch, selection of meas. range,
min- and max-value memory,
hold-function calibration etc.
Power supply:
9V battery, type IEC 6 F 22 (included)
as well as additional d.c. adapter (internal pin diameter 1.9
mm)
for external 10-12 V direct voltage supply (suitable power
supply GNG 10/3000)
Power-Off-Function:
Device will be automaticially
switched off if no key is pressed/
no interface communication takes
place for the time of the power-off delay.
The power-off delay can be set to
values between 1 and 120 min. or deactivated.
Low battery warning:
'bAt'
Power consumption:
approx. 1.5 mA
Dimensions:
142 x 71 x 26 mm (H x W x D)
impact-resistant ABS plastic housing, membrane keyboard, transparent
panel.
Front side IP65, integrated pop-up
clip for table top or suspended use
Weight:
approx. 160 g (cpl. with battery)
Temperature compensation:
automatically
via temperature sensor integrated in probe housing
Compensation of air pressure:
the O2 -concentration will be
compensated according to the abs. atmospheric pressure set (500 to 2000 hPa)
Calibration:
1-point calibration: extremely simple quick
(calibration in atmospheric air (press button to compensate unit to 20.9
%)
2-point calibration:
first point at atmospheric air (20.9 %) and second point freely selectable
Application:
Wide range of application
for your home, job and hobby, e.g.:- Bio chemistry: Oxygen monitoring an breeding
chambers for cell cultures.
Monitoring of fermenting process fo fruits infermentation plants etc.
- Medicine: Monitoring of oxygen concentration in
respirators; checking of breathing, gmh369
monitoring of oxygen concentration in incubators, oxygen tents etc.
- Food technology: Monitoring of residual
oxygen in packages (e.g. coffee, tea, etc.).
Monitoring of oxygen content during production processes
- Safety technology, safety at work: Oxygen
monitoring in mines/pits, underground parking lots,
wine cellars,
cooling chambers, greenhouses or stores.
Oxygen
monitoring or alarm in case of danger of suffocation when working in tanks,wells
etc.
- Air conditioning and ventilation technolgy:
Oxygen measurements, air quality monitoring,
measuring of
oxygen concentration in enclosed air conditioning systems, etc.
- Sports: Checking of oxygen conent in compressed
air breathing apparetuses (diving, etc.)
oxygen
monitoring for gliding.
The device can only be used to check within these applications.
-> no substitute for approved monitoring device!
Suitable
electrodes (GGO 369 / GOO 369)
EBS9M
Software
additional
accessories
Air oxygen measuring-device with double display for oxygen and temperature, delivery WITHOUT probe and accessories
Additional features of the GMH 3691 GL:
O2-Partial pressure measuring
compensation of atmospheric pressure by entering the atmospheric pressure
1-point or 2-point-calibration
Please order separately
certificate of calibration
Software EBS 9M and further
accessories
Please order furthermore separately:
= data sheet as *.pdf
= instruction manual as *.pdf
deutsche Beschreibung
Your article: GMH 3691 GL;
Article No. GMH 3691 GL.
The total price of the article is: EUR
The net price of the article GMH 3691 GL
is: EUR 147.9