Specification:
|
|
| Meas.
range: |
0.00 to 14.00 pH |
| Accuracy: |
0.02 pH + 1 Digit (at nominal temperature =25°C) |
| Output signal: |
4-20 mA (2-wire- standard),
0-10 V (3-wire - optionally) |
|
Electric isolation: |
Input electrically isolated |
|
Auxiliary energy: |
12 ... 30 V DC at option 0- 10 V: 18 ... 30 V
DC |
| permissible impedance: |
(at 4-20mA): RA
(Ohm) = UV (V) - 14 / 0,02 A) |
| permissible burden: |
(bei 0 - 10 V): RL > 3000 Ohm |
| Electrode: |
any pH-electrode
of our program is suitable (electrode is not included in
scope of supply!) |
| input resistance: |
1012
Ohm |
| electrode socket: |
Cinch-Socket
(Standard)
Optionally available with BNC-
and
DIN-socket (please specify on your order) |
|
Temperature compensation: |
0 ... 100
°C,
manually via 3 keys or
automatically vie external Pt1000 sensor |
| calibration: |
Calibration: via 3 keys and integrated LCD |
| Temperature
sensor socket: |
3.5 mm phono plug or banana plug for the
DIN-socket type |
|
Display: |
approx. 10 mm high, 3 1/2 digit display |
| Working temperature meas. transducer: |
0 ...
50 °C |
|
Storage temperature: |
- 20 ... +70 °C |
| Electric
connection: |
elbow-type plug according to DIN 43650 (IP 65),
maximal wire cross section:
1,5 mm², wire diameter from 4.5 to 7mm
wire diameter from 4.5 to 7
mm |
|
Housing: |
ABS (IP65) with the exception of electrode and temperature
connection sockets (complete with IP 65 upon request) |
|
Dimensions: |
82 x 80 x 55 mm (W x H x
D) |
| Mounting: |
with fixing holes for wall mounting (accessible after removal of
cover)
Mounting distance: 70 x
50 mm (B x H)
Fixing screws:
max. schaft-diameter 4 mm |
| |
|
|
Order example: |
|
| GPHU 014 MP / BNC, AV010: |
GPHU 014 MP with BNC-electrode socket and 0-10V output signal |
| |
|
| |
|
|
Optionen: |
|
| AV 010: |
output signal 0 - 10 V |
| BNC: |
electrode socket: BNC |
| DIN: |
Electrode socket: DIN |
| MB: |
limited measuring range (please state the desired range) |
| |
|
|
Accessories: |
all Pt1000-probes |
| GTF 2000 WD |
Pt1000-temperature probe, water proof type |
| GE 100 |
Standard electrode, Cinch-plug |
| GE 107 |
special-pH-electrode with integrated Pt1000-temperature
sensor, DIN-Stecker |
| PG 13,5 Gewinde |
plug on thread adapter for pressureless use |
|
GAK 1400 |
working and calibration set |
| |
|